Online Service Report — Sample
A live example of the secure report portal every Audiology 4 Vets client receives. In production, each client logs in to see only their own facility’s records, updated automatically from the field.
Client / Facility
Sample Regional Medical Center
Report #
SR-2026-04871
Service Date
Oct 4, 2026
Technician
D. Lockett, CBET
Equipment Serviced
| Asset ID | Device | Mfr | Model | Location | Result | Next PM |
|---|---|---|---|---|---|---|
| BMX-10442 | Anesthesia Workstation | GE / Datex | Aisys CS2 | OR-3 | PASS | 2026-10-04 |
| BMX-10443 | Patient Monitor | Philips | IntelliVue MX750 | OR-3 | PASS | 2026-10-04 |
| BMX-10461 | Infusion Pump | BD | Alaris 8100 | ICU-2 | FOLLOW-UP | 2026-07-22 |
| BMX-10480 | Defibrillator | ZOLL | R Series | ER | PASS | 2026-10-05 |
Service Performed
- Visual inspection & cleaning completed
- Electrical safety test (IEC 62353) — leakage within limits
- Performance verification to manufacturer spec
- Alarms & safety systems verified
- Calibration confirmed / adjusted
- Firmware at current approved revision
Field Notes
Alaris 8100 (BMX-10461) flagged an intermittent occlusion alarm; door mechanism cleaned and retested, monitoring for recurrence — follow-up 07/22. All other assets passed and returned to service.
Compliance: Electrical-safety and performance records logged to the client portal and archived for Joint Commission / CMS audit readiness.
Updated automatically from the field
BiomedRx’s AI-powered field app writes each result straight into this portal as the technician completes the work — no manual data entry, no lag. Clients see status the moment a device returns to service.
Devin Lockett
Certified Biomedical Equipment Technician · Audiology 4 Vets
Digitally signed · Oct 4, 2026 4:12 PM
Report ID SR-2026-04871